It is anticipated that the Bidder will collaborate closely with PATH and goal Hospitals for the waste handling

building/area and will make available all needed technical information, drawings, etc.,

 

The following waste streams will likely be treated in the newest HCW treatment centre (s):

 

Infectious waste

 

Sharps waste

 

Pathological waste

 

Pharmaceutical waste containing cytotoxic or genotoxic waste

 

Small amounts of chemical waste

TECHNICAL SPECIFICATION FOR PROCUREMENT OF”MEDICAL WASTE TREATMENT PLANT”

Lot-A

Specification

1. Hydroclave for Medical waste Therapy Plant: Quantity: 02 Nos.

General Prerequisites

The processor must be approved to sterilize  & see to these kinds of medical waste:

-Liquids: Blood, body fluids .

-Blood bags & tubing, gloves, Gowns, curtains, etc..

-Contaminated packaging: Lotion etc..

-Sharps & Sharp containers

-Disposable instruments

-Glass products

-Human and animal tissue (surgical and pathologic) –Based upon MWM regulations

-Cultures, stocks and glassware

The processor must comply with a documented efficiency level of 6log10 reduction in spore survival (99,9999%) of their

Bacillus Stearothermophilus

The processor should reduce the volume of the waste by 60% to 80% through complete dehydration and sterilization of

the waste

The  processor must have ZERO harmful emission level and meet or exceed all criteria demanded by the EU.

The whole waste treatment system needs to be offered as a complete unit consisting of subsequent gear and

ancillaries-

(A)The sterilization processing unit

(B)Steam Boiler

(C)Tub washer

(D)Heavy duty outside shredder-

(E)Conveyer belt-based on Manufacturer’s Recommendations

(F)Bin Lift/Tipper

(G)Spore tester

(A)Features and Specifications of the Sterilization Processor unit:

Technical specification of the processor:

Steam autoclave in conjunction with a heavy duty Fragmenting System within the vessel/chamber

The vessel must be of Double Chamber

Capacity (kg/load): minimum 200 kg/batch/cycle

Volume of 140-190 liters of bulk each cycle

Average Cycle time: 60 minutes

Running temperature: 121◦C to 135◦C

Running pressure: 1.1 Bar to 2.4 Bar